  ima 1000 watts power supply series for medical and industrial applications product data sheet
1000 watts power supply series for medical and industrial applications model variants model number 1) input voltage range main dc output auxiliary dc output remote on/off standard setting 2) ac (vac) dc (vdc) voltage (vdc) current (a) voltage (vdc) current (a) ima-x1000-12-yypli 80 to 275 120 to 300 12 84 5 2 off ima-x1000-12-yyply on IMA-X1000-24-YYPLI 24 42 off ima-x1000-24-yyply on ima-x1000-48-yypli 48 21 off ima-x1000-48-yyply on 1) ima-x1000: x = s for standard version (e.g. ima-s1000-24-yyply), x = c for conformal coated version (e.g. ima-c1000-24-yyply) 2) model yypli and yyply have different settings for remote on/off, see other features, p. 4. ac/dc input (j1) ima-x1000-12 ima-x1000-24 ima-x1000-48 nominal input voltage 100 vac to 240 vac ac operating input voltage range 80 vac to 275 vac nominal input frequency 50 / 60 hz input frequency range 47 hz to 63 hz dc input voltage range 120 vdc to 300 vdc maximum input current 15 a at 80 vac / 9.5 a at 120 vdc effciency @ 70% load 1) see fig. 12 to fig. 14 @ 230 vac 93% 94% 94% @ 115 vac 91% 92% 91.5% max inrush current 2) < 20 a input fuse dc input compliant, dual 16 a fuses used power factor 3) 0.9 (typical) 1) excluding fan power 2) hot and cold turn on 3) en 61000-3-2, class a compliant features safety rated for medical, industrial and it wide operating input voltage range: 80 vac to 275 vac or 120 vdc to 300 vdc wide adjustable output voltage range (+/- 20%) 5 vdc standby output high effciency: up to 94% size: 5 x 8.25 x 1.6 in (1u design) variable speed fan control low acoustic noise level of less than 39 db(a) active current sharing 2 mopp pmbus? compatible for control, programming and monitoring 500,000 hour mtbf optional conformal coating 3 years warranty data sheet ima x1000 power supply series v2.0 2016-08-09 2 ima-x1000 power supply series
main dc output (j2) ima-x1000-12 ima-x1000-24 ima-x1000-48 nominal output voltage 12 v 24 v 48 v output voltage adjustment range 9.6 v to 14.4 v 19.2 v to 28.8 v 38.4 v to 56.0 v maximum output power 1,000 w output voltage regulation total 2.25% over line full input range, full load 0.25 % over load nominal input, full load range 1% over temperature nominal input, full load, full temperature range 1% maximum output current 84 a 42 a 21 a maximum output capacitive load 10,000 f dynamic load regulation 1) < 5% pard (20 mhz) 2) < 120 mv < 150 mv < 200 mv turn on overshoot < 2% output rise time < 100 ms hold up time 20 msec nominal start up time ac off --> on nominal input, max. load < 2.5 s remote off --> on nominal input, max. load < 150 ms output over voltage protection yes, latch mode 15 v to 17.5 v 30 v to 35 v 58 v to 65 v output over current protection yes, at 108% to 140% of maximal output current; auto recovery short circuit protection yes, auto recovery over temperature protection yes, auto recovery remote sense 3) total voltage drop com - pensation for +v_sense and -v_sense connec - tions (j3 pins 13 and 14) to the output load 200 mv 1) 50% step from 5% load,1 a/s, 10 f tan and 1f ceramic capacitor 2) 10 f tan and 1f ceramic capacitor 3) do not short or reversely connect +v_sense and -v_sense. doing this can cause damage to the power supply. auxiliary dc output (j3) ima-x1000-xx connector type molex, part number 87833-1420, 14 pin, see fig. 15, page 11 nominal output voltage 5 v output voltage adjustment range C output voltage regulation total 2.25% over line full input range, full load 0.25% over load nominal input, full load range 1% over temperature nominal input, full load, full temperature range 1% maximum output current 2 a maximum output capacitive load 1,000 f output over voltage protection yes, at 5.7 v to 6.5 v; latch mode output over current protection yes, at 108% to 140% of maximal output current; auto recovery short circuit protection yes, auto recovery over temperature protection yes, auto recovery 3 data sheet ima x1000 power supply series v2.0 2016-08-09 ima-x1000 power supply series
galvanic isolation ima-x1000-xx input to output reinforced 4000 vac; 2 x mopp input to case basic 1500 vac; 1 x mopp output to case basic 1500 vac; 1 x mopp leakage currents ima-x1000-xx ac leakage current from input to earth ground measured at mains voltage at 60 hz at 63 hz normal condition (low line) 132 vac < 150 a < 150 a single fault condition (low line) 132 vac < 250 a < 260 a normal condition (high line) 264 vac < 300 a < 300 a single fault condition (high line) 264 vac < 500 a < 520 a ac leakage current from output to earth ground measured at mains voltage typical at 60 hz 1) maximum value at 63 hz 1) limit per iec 60601-1 normal condition (low line) 264 vac 55 a < 70 a 100 a single fault condition (low line) 264 vac 43 a < 80 a 500 a normal condition (high line) 264 vac 172 a < 230 a 500 a single fault condition (high line) 264 vac < 1250 a < 1800 a 5000 a 1) meets iec 60601-1 bf leakage current limit other features ima-x1000-xx current share bus pin j3 pin 11 (current share_v) voltage at cs pin will vary linearly with load current on main output, and will be 6 v at rated load current, when the output voltage is at its rated value. power good pin j3 pin 9 (pwr_good) open collector. as soon as ac input voltage and dc output voltage are in the predefned range, the pwr_good signal is set to high. green led will turn on as soon as pwr_good signal is set to high component derating guideline refer to ipc 9592b and to delta guideline or-ing redundant operation with active circuit sharing, see application note redundant operation, p. 10 sda, scl for i 2 c internal 10 k pull-up resistor to internal 3.3 v ima-x1000-xx-yypli ima-x1000-xx-yyply remote on/off pin 1) j3 pin 10 (remote on/off) remote on/off (j3 pin 10) and 5vsb_rtn (j3 pin 3 or j3 pin 4 or j3 pin 7) main dc output remote on/off (j3 pin 10) and 5vsb_rtn (j3 pin 3 or j3 pin 4 or j3 pin 7) main dc output shorted off shorted on open on open off 1) logic can be switched with pmbus? data sheet ima x1000 power supply series v2.0 2016-08-09 4 ima-x1000 power supply series
environmental conditions ima-x1000-12 ima-x1000-24 ima-x1000-48 ambient operating temperature range 1) (see fig. 8, page 10) standard mounting orientation (see fig. 1, page 8 ) --20 c ... +70 c (-4f to +158 f) other mounting orientations -20 to +65 c (-4 to +149 f) ambient storage temperature range -40 c ... +85 c (-40 f to +185 f) output power derating versus input voltage when ac input voltage is < 90 vac, the output power will be reduced by 20 w per 1 v. (see fig. 7, page 10) versus ambient temperature (see fig. 8, page 10) standard mounting orientation (see fig. 1, page 8) when ambient temperature is > 50 c (122 f), the output power will be reduced by 25 w per 1 c. other mounting orientations when ambient temperature is > 45 c (113 f), the output power will be reduced by 25 w per 1 c. output current derating versus output voltage when output voltage is > 12 vdc, the output current is reduced by 6.08 a per 1 v (see fig. 9, page 10). when output voltage is > 24 vdc, the output cur- rent is reduced by 1.52 a per 1 v (see fig. 10, page 10) . when output voltage is > 48 vdc, the output cur- rent is reduced by 0.4 a per 1 v (see fig. 11, page 10) . relative humidity < 95% (non-condensing) operating altitude 1) 2) -200 m to 5,000 m (-650 ft to 16,400 ft) shock test (non-operating) iec 60068-2-27 compliant, 50 g, 11 ms, 3 shocks for each direction vibration iec 60068-2-6 compliant, 2.09 grms, 5 - 500 hz, 20 minutes per side (3 planes) pollution degree 2 1) ambient operating temperature decreases by 1 c per 305 m (1000 ft) altitude increase 2) maximum operating altitude requirements for different types of products, see safety standards and directives 1), p. 6 reliability ima-x1000-xx cmtbf 1) 500,000 hours expected capacitor life time 2) 10 years fan l 10 life @ 40 c 70,000 hours warranty 3 years 1) telecordia sr-332, issue 2, 25 c, 90% confdence level 2) nominal input voltage, 45 c (113 f), 80% load 5 data sheet ima x1000 power supply series v2.0 2016-08-09 ima-x1000 power supply series
emc ima-x1000-xx applied standards criteria radiated emissions 1) en 55011, en 55022 and fcc, class b conducted emissions 1) en 55011, en 55022 and fcc, class b power line harmonics en 61000-3-2, class a voltage ficker en 61000-3-3 esd en 61000-4-2, level 4, 8 kv contact, 15 kv air a radiated immunity en 61000-4-3, level 2, 3 v/m a electrical fast transient en 61000-4-4, level 3, 2 kv a surge immunity en 61000-4-5, level 3, 1 kv dm, 2 kv cm a conducted rf immunity en 61000-4-6, level 2, 3 vrms a power frequency magnetic feld en 61000-4-8, level 2, 3 a/m a voltage dips and sags en 61000-4-11, 30%, 500 ms en 61000-4-11, 60%, 100 ms en 61000-4-11,100%, 10 ms en 60601-1-2, 30%, 500 ms en 60601-1-2, 60%, 100 ms en 60601-1-2, 100%, 10 ms en 60601-1-2, 100%, 5000 ms a b a a b a b ring wave en 61000-4-12, level 3, 1 kv dm, 2 kv cm a voltage fuctuations en 61000-4-14, class 3 a 1) power supply unit inside a dummy system safety standards and directives 1) ima-x1000-xx iec/en 60950-1, edition 2 and all national devia - tions ul 60950-1/csa 22.2 no 60950-1, edition 2; 5000 m (16,400 ft) altitude, 120 v to 300 vdc and 100 v to 240 10% vac (ul file e191395) iec/en 60601-1, edition 3 (tested against edition 2, too) and all national deviations iec 60601-1(2005), en60601-1(2006) ansi/aami es 60601-1(2005) can/csa c22.2 no. 60601-1 (2008); 3,000 m (9,800 ft) altitude, 100 v to 240 vac 10% (ul file e325662) protection class i 1) designed to support type b applied part end product requirements ecological characteristics ima-x1000-xx waste electrical and electronic equipment directive (weee) 2002/96/ec rohs - eu directive 2011/65/ec rohs compliancy data sheet ima x1000 power supply series v2.0 2016-08-09 6 ima-x1000 power supply series
mechanical data ima-x1000-xx dimensions (l x w x d) 209.5 x 127 x 40 mm (8.035 x 5 x 1.57 in) weight 1.6 kg (3.53 lb) indicator green led cooling system 1) 2 fans with variable speed control ac/dc input terminal block block m3.5 x 3 pins main dc output terminal block block m5 x 2 pins auxiliary dc output + signals port connector x 14 pins acoustic noise 1) 2) < 39 db(a) 1) to keep the noise low the fan will be turned off in standby mode 2) at 1 hz to 20 khz and a distance of 1 m. test conditions: 100 vac, 100% load, ambient temperature 30 c (86 f) options model main output voltage standby output leakage current main output adjustable open frame u channel enclosed convection cooling fan fan, airfow from end to front fan, airfow from front to end top fan solution active current sharing remote on/off coated 1) ima-s1000-12v 12 v 5 v/2 a 300 a - ima-s1000-24v 24 v 5 v/2 a 300 a - ima-s1000-48v 48 v 5 v/2 a 300 a - ima-c1000-12v 12 v 5 v/2 a 300 a ima-c1000-24v 24 v 5 v/2 a 300 a ima-c1000-48v 48 v 5 v/2 a 300 a included on request - not available 7 data sheet ima x1000 power supply series v2.0 2016-08-09 ima-x1000 power supply series
mounting orientations fan side connector side 20 mm (0.79 in) fig. 1: standard mounting orientation fan side connector side fig. 2: vertical mounting bottom to p fig. 3: mounting on the left side bottom to p fig. 4: mounting on the right side mounting holes fan side connector side side view mounting holes mounting holes bottom view fig. 5: position of mounting holes data sheet ima x1000 power suppl series v2.0 2016-08-09 8 ima-x1000 power supply series
dimensional drawings 127.0 mm +0.5 -0.5 (5.0 in ) +0.02 -0.02 209.5 mm +1.0 -1.0 (8.25 in ) +0.04 -0.04 airflow led vr j2 j1 j3 pin2 pin1 145.0 mm (5.71 in) 30.5 mm (1.2 in) 20.0 mm (0.79 in) 40.6 mm +0.5 -0.5 (1.60 in ) +0.02 -0.02 m3 screw hole (2x) both sides 160.0 mm (6.30 in) 15.5 mm (0.61 in) 7.5 mm (0.30 in) m3 screw hole (4x) bottom side 112.0 mm (4.41 in) fig. 6: dimensional drawing ima-x1000-xx notes: base plate mounting, m3 thread holes, maximum penetration 4.0 mm (0.16 in) from outside face of chassis, maximum torque 0.6 nm (5.31 lb-in) (j1) input terminal block, switchlab t14-emii03, m3.5 screw in 3 positions, torque 1.3 nm (11.5 lb-in) (j2) output terminal block, dinkle 0166-8002c, m5 screw in 2 positions, torque 2.4 nm (21.24 lb-in) (j3) mating connector for j3 is either molex, part number 51110-1450 (without locking ramp), or molex part number 51110-1451 (with locking ramp). the connector is not shipped with the power supply unit. 9 data sheet ima x1000 power supply series v2.0 2016-08-09 ima-x1000 power supply series
curves ima-x1000-xx ima-x1000-xx 100 80 80 90 275 output power [%] input voltage [vac] fig. 7: output power versus input voltage 100 50 -20 50 70 [c] -4 122 158 [f] 45 65 149 113 standard mounting orientation all other mounting orientations output power [%] ambient temperature tested at 90 vac input fig. 8: output power versus ambient temperature ima-x1000-12 ima-x1000-24 output current [a] 84 69.4 output voltage [vdc] 9.6 12 14.4 fig. 9: output current versus output voltage 12 v output current [a] 42 34.7 output voltage [vdc] 19.2 24 28.8 fig. 10: output current versus output voltage 24 v ima-x1000-48 output current (a) 21 17.8 output voltage (vdc) 38.4 48 56 fig. 11: output current versus output voltage 48 v data sheet ima x1000 power supply series v2.0 2016-08-09 10 ima-x1000 power supply series
curves (continued) ima-x1000-12 ima-x1000-24 efficiency [%] 94 96 92 90 88 86 84 82 80 78 output power [%] 20 40 60 80 100 230 vac 115 vac fig. 12: typical effciency curves 12 v efficiency [%] 94 96 92 90 88 86 84 82 80 78 output power [%] 20 40 60 80 100 230 vac 115 vac fig. 13: typical effciency curves 24 v ima-x1000-48 efficiency [%] 94 96 92 90 88 86 84 82 80 78 output power [%] 20 40 60 80 100 230 vac 115 vac fig. 14: typical effciency curves 48 v pin assignment (j3) ima-x1000-xx 1 2 3 5 7 9 11 13 6 8 10 12 14 4 fig. 15: pin assignment j3 terminal block pin assignment pin assignment 1 +5vsb 2 +5vsb 3 5vsb_rtn 4 5vsb_rtn 5 scl 6 sda 7 5vsb_rtn 8 5vsb 9 pwr_good 10 remote on/off 11 current_share_v 12 address 13 +v_sense 14 -v_sense mating connector type: molex, part number 51110-145x 11 data sheet ima x1000 power supply series v2.0 2016-08-09 ima-x1000 power supply series
circuit diagrams ima-x1000-xx ima-x1000-xx 1 k +5vsb power supply unit 5vsb_rtn remote on/off j3 fig. 16: circuit diagram j3 pin 10 (remote on/off) 1 k 3.3v_c power supply unit 5vsb_rtn pwr_good j3 fig. 17: circuit diagram j3 pin 9 (pwr_good) ima-x1000-xx pwr_good main dc output ac input 10 ms 10 ms 20 ms fig. 18: power good function timing 1) )rrsrodj1rpdorsrodjoogfdrsrodj v1rpdorsrodj 1) data sheet ima x1000 power suppl series v2.0 2016-08-09 12 ima-x1000 power supply series
this page is intentially left empty 13 data sheet ima x1000 power supply series v2.0 2016-08-09 ima-x1000 power supply series
data sheet ima x1000 power supply series v2.0 2016-08-09 all information and specifcations are subject to change without prior notice sales contact europe delta electronics (netherlands) bv zandsteen 15 2132 mz hoofddorp the netherlands deu.sales@deltaww.com www.deltaenergysystems.com usa delta products corporation 46101 fremont blvd. fremont, ca 94538 usa na.sales@deltaww.com www.deltaenergysystems.com other regions delta energy systems (germany) gmbh tscheulinstrasse 21 79331 teningen germany im.sales@deltaww.com www.deltaenergysystems.com


▲Up To Search▲   

 
Price & Availability of IMA-X1000-24-YYPLI

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X